CL System
Affordable Technology

DSG04 (20Msps version) - Technical Specifications

Scope

Maximum Sampling Rate 20 Millions Samples per second
Minimum Sampling Rate 2000 Samples per second
Maximum sample lengths of 32 seconds with 64K buffer @ 2000sps
Buffer Size Selectable 2K, 4K, 8K, 16K, 32K & 64K 12bit data points

Analogue

Analogue channels input Ranges +-1.0V and +-10V (+-100V with X10 probes)
Analogue input modes Software selectable AC or DC for analog inputs.
Analogue input bandwidth ~10MHz
Analogue input impedance ~ 1 Mohms // 22pf
Maximum allowable voltage at input +-20V
Analogue input resolution 12-bit resolving to less than 500uV per bit

Digital

Digital channels input Approximately <0.5V for logic 0 and >2 V for logic 1
Digital input mode DC only
Digital input bandwidth ~ 100MHz
Digital input impedance ~ 20K // 22pf
Maximum allowable voltage range at input more than -2V and less than +10V
Digital input resolution 1-bit.  Logic Low or High

Trigger

Trigger modes Instant or Digitally Triggered (with filter option) with +/- slopes
Trigger source Channel A, B, C or D
Pretrigger options OFF, 10 or 100 sample points

Display

Display grid

25mV/div to 2.5V/div (or 25V/div with X10 probe)

Input

Input connectors All standard BNC
Stability (@ 20 oC) Zero reading: +- 4 bits, Noise: ~ +- 2 bits at 2Msps (with inputs grounded)

Spectrum Analyzer

Type FFT (Fast Fourier Transform)
Range 0 to 10 MHz

Signal Generator

Frequency Range 5Hz to 500KHz settable to 1 percent. Accuracy to within 3%
Built-in signal types Sine, Triangular, Saw, Square wave
Built-in signal modes Normal or Sweep
Sweep Ratios of X2, X5 X10 and X20
Sweep start freq.

5,10,50,100,500,1000,5000,10000,50000,100000 Hz

Custom waves Selectable 2K, 20K, 200K, 2M & 20M Hz update rates
Custom wave loaded via simple text file, raw data or Excel CSV file
Custom value of 4095 = 2V  & 0 = 0V (12bits resolution)
Signal Amplitude Built-in waveforms 0-1V and 0-2V pp
Output impedance 200 Ohms

Counter

Count Modes Frequency or Counter
Minimum signal pulse width ~ 10nS
Frequency count range 1 to 100,000,000 Hz 9-digit display with 1Hz resolution
Frequency accuracy ~ 0.01% +- 1 digit, stability +-100ppm/oC
Frequency source inputs Analogue channels A or B or digital channels C or D
Counter range 0 to 1,073,741,823 counts
Counter source Analogue channels A or B or digital channels C or D or 1MHz reference
Counter control Ch C (+edge start &ndash;edge stop), D (-edge start +edge stop) or Manual
Pulse width measurement (+ or -) 1 us to 1000 seconds using internal 1MHz reference

Most counter functions are also available in Generator Mode

Scope Modes

SCOPE, GENERATOR, SCOPE & GENERATOR, FFT, COUNTER

© Copyright CL SYSTEM 2005